Cornelius DPC 230 Specifications Page 31

  • Download
  • Add to my manuals
  • Print
  • Page
    / 70
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 30
Fluorescence Correlation 25
A typical optical setup is shown in Fig. 34. A CW laser beam is focused into the sample
through a microscope objective lens. The fluorescence light from the sample is collected by
the same lens, separated from the laser by a dichroic mirror, and fed through a pinhole in the
upper image plane of the microscope lens. Light from above or below the focal plane is not
focused into the pinhole and therefore gets substantially suppressed. With a high-aperture ob-
jective lens the effective sample volume is of the order of a femtoliter, with a depth of about
1.5 µm and a width of about 400 nm. A similarly small sample volume can be obtained by
two-photon excitation. Typical FCS instruments use several detectors in different wavelength
intervals or under different polarisation. Due to the similarities in the optical system, FCS is
possible with most confocal laser scanning microscopes [2].
Polarising
beamsplitter
Dichroic
Dichroic
Dichroic
Microscope
Lens
Laser
Detectors
Detectors
Pinhole
CW
Fig. 34: Optical setup for FCS measurements
For good FCS results high detection efficiency is essential. The reason is the unusual signal-
to-noise behaviour of correlation experiments. In a fluorescence lifetime or FLIM measure-
ment a 50% loss in efficiency can be compensated by increasing the acquisition time by a fac-
tor of two. In an FCS measurement a 50% loss in detection efficiency requires the acquisition
time to be increased by a factor of four to obtain the same signal-to-noise ratio. The detectors
of FCS instruments are therefore single-photon avalanche photodiodes (SPADs) or high-
efficiency PMTs with GaAsP cathodes [2].
The electronic setup of an FCS system is shown in Fig. 35. SPADs are connected via a cable
adapter (see Fig. 20) to the LVTTL Inputs of the DPC. Any SPAD module with TTL or
LVTTL output can be used. PMTs are connected to the CFD inputs, see Fig. 35, right.
DPC-230 module
LVTTL
Inputs
Extension cables
SPADs
id Quantique, MPD,
Perkin Elemer, Sensl
Detectors
DPC-230 module
H7422 Power Supply and
DCC-100 Detector Controller
Overload Shutdown
HFAC-26-2
Preamplifier
H7422P-40
H7422P-40
Fig. 35: FCS setup with DPC-230. Left: SPAD detectors. Right: GaAsP PMT modules
The measurement control part of the System parameters is shown in Fig. 36. ‘Operation
Mode’ is set to ‘Absolute Time’, ‘Correlation’. The settings shown on the left calculate and
display FCS curves, the accumulated number of counts, and an MCS (intensity) trace online.
The time-tag data are not saved. The settings shown on the right calculate the same curves, but
simultaneously save the time-tag data into a file ‘fluorescein-a.spc’. The maximum file size is
1000 Mb. The measurement stops when this file size is reached. The measurement can also be
Page view 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 69 70

Comments to this Manuals

No comments